8 Cleaning and disinfection
Result
8.3.6 Steam sterilizing parts (optional)
Requirement
Result
116
EN
MEDUMAT Easy
The parts are machine cleaned and disinfected.
Clean and disinfect the hose protection sleeve.
1. Open the hose protection sleeve completely.
2. Wash the hose protection sleeve in the domestic or industrial
washing machine at 60°C adding the cleaning agent and
disinfectant specified in the cleaning and disinfection plan
"8.2 Cleaning and disinfection plan", page
When doing so, note: Compliance with the manufacturer's
specifications must be assured.
3. Allow the hose protection sleeve to dry completely.
If steam sterilization is to be performed, do so in accordance with
the internal processes in place at your facility.
The parts intended for steam sterilization are visibly clean.
The parts intended for steam sterilization have been
disinfected.
1. The parts approved for steam sterilization can be found in the
cleaning and disinfection plan
disinfection plan", page
2. Steam sterilize parts with a device corresponding to EN 285.
When doing so, note:
At 134°C with a hold time of 5 minutes.
or
At 132°C with a hold time of 4 minutes.
The sterilizer manufacturer's instructions must be observed
and followed.
The parts are steam sterilized.
CPR
(see "8.2 Cleaning and
100).
(see
100).