Table of Contents

Advertisement

3. Insert spontaneous respiration insert (2) in spontaneous
respiration side (4) of the reusable patient valve.
4. Insert new disk diaphragm (5) in expiration side (6) of the
reusable patient valve.
5. Insert a new lip membrane (9).
When doing so, pay attention to the insertion direction symbol
on the reusable patient valve
reusable hose system with reusable patient valve", page
6.
Warning! Treatment fault due to patient valve lip
membrane installation errors!
Check the correct positioning of the lip membrane in
accordance with the installation direction symbol on the
reusable patient valve
7. Screw the connection for the ventilation hose (7) onto the
reusable patient valve lid (8) in a clockwise direction.
8. Install the reusable patient valve lid (8) in a clockwise direction.
9. Assemble the reusable hose system
reusable hose system", page
10. Perform a function check
Result
The membranes in the reusable patient valve have been changed.
10 Maintenance
(10)
(see "3.6.4 Labels on the
(10) .
(see "4.5 Assemble
46).
(see "5 Function check", page
CPR
MEDUMAT Easy
33).
51).
EN
129