8 Cleaning and disinfection
8.3.1 Preparing cleaning and disinfection
Requirement
106
EN
MEDUMAT Easy
Risk of infection due to inadequate hygienic safety!
Adequate cleaning and disinfection is not guaranteed if the
mounting plate with fastening strap is not removed and the belt
around the buckle is not slackened prior to the cleaning and
disinfection. If contaminated surfaces come into contact with the
skin of the patient or user, this can lead to infections. This can
result in injury to the patient or user.
 Remove the mounting plate and fastening strap from the
device before cleaning and disinfection.
 Slacken the buckle on the fastening strap.
The device is switched off
page
84).
The accessories are removed
The compressed gas tube is assembled on the device.
1.
Warning! Risk of injury and damage to the device
caused by ingress of liquids!
Do not remove the compressed gas hose from the compressed
gas connection.
2.
CAUTION! Risk of injury and damage to the device
caused by ingress of liquids!
Turn the connection for ventilation hose downward.
CPR
(see "6.6 Switching the device off",
(see "7 Disassembly", page
89).