5 Function check
Result
60
EN
MEDUMAT Easy
4.
CAUTION! Risk of injury from improperly removed
testing bag!
Grasp the testing bag by the connector and pull the bag and
connector off the patient valve.
Check whether the device triggers the pAW
alarm after the second inspiration breath.
5. Connect the testing bag to the inspiration side of the patient
valve again.
The tidal volume and airway pressure measurement have been
checked.