6 Operation
Requirement
78
EN
MEDUMAT Easy
Switching on CPR mode
The device is ready for use
page
64).
MEDUtrigger is connected
hose system and MEDUtrigger to the device", page
A tidal volume is set with the adjusting knob for the ventilation
parameters
(see "6.2.1 Setting the respiratory rate and tidal
volume", page
The device is switched on
page
72).
Voice prompt and/or the metronome is switched on
Activating/deactivating the voice prompt", page 85)
"6.8 Activating/deactivating the metronome", page
Risk of limited treatment due to insufficient patient
monitoring!
If the metronome is switched on, the audible alarms in CPR mode
pause during the chest compression and voice prompt phases. In
case of limited patient monitoring, this may compromise the
treatment. This can injure the patient.
 Monitor patients in CPR mode continuously.
Limited patient monitoring via deactivated voice prompt for
the pAW
and pAW
Voice prompt is deactivated for the pAW
alarms in CPR mode.
In case of limited patient monitoring, this may result in injuries to
the patient.
 Monitor patients in CPR mode continuously.
CPR
(see "6.1 Preparing for ventilation",
(see "4.6 Connecting the patient
70).
(see "6.3 Switching the device on",
/
Apnea alarms in CPR mode!
48).
(see "6.7
and
(see
87).
/
and pAW
Apnea