Weinmann MEDUMAT Easy CPR Instructions For Use Manual page 65

Ventilators
Table of Contents

Advertisement

2. Open the oxygen cylinder slowly.
The contents gauge (1) displays the oxygen cylinder pressure.
3. Calculate the remaining operating time to ensure that the
device does not stop unexpectedly
page
141).
4. Connect accessories.
Connect ventilation mask or tube
the ventilation mask or tube", page 65)
Connect breathing system filter
breathing system filter", page 67)
Connect PEEP valve
valve", page 69)
Result
The device is ready for use.
6.1.1 Connecting the ventilation mask or tube
Delayed treatment due to incorrect position of MEDUtrigger
on the patient valve!
If MEDUtrigger is incorrectly connected to the patient valve, it will
not be possible to properly attach the ventilation mask, which can
lead to incorrect or delayed treatment. This can injure the patient.
 Push MEDUtrigger fully onto the inspiration side of the patient
valve.
1
(see "6.1.3 Connecting the PEEP
(see "14 Appendix",
(see "6.1.1 Connecting
(see "6.1.2 Connecting the
CPR
MEDUMAT Easy
6 Operation
EN
65