6 Operation
6.4.4 Monitoring the patient
82
EN
MEDUMAT Easy
Risk of incorrect treatment due to insufficient patient
monitoring!
If the patient and device are not supervised and monitored during
ventilation, delayed responses of medical personnel to alarms and
error messages may result in serious or life-threatening injuries to
the patient and to incorrect treatment.
 The patient and device must be kept under continuous
observation and monitoring during ventilation.
 Use additional external monitoring during ventilation (e.g.,
SpO
or etCO
).
2
2
1. Check the ventilation pressure. To do this, read off the
ventilation pressure on the ventilation pressure display.
2. Check the ventilation parameters.
3. Check the cause of the emitted alarms
messages", page
High air resistances, e.g., due to obstructions of the airways or
during external chest compression, change the set tidal volume.
With reduced lung compliance, the device reacts as per the
diagram shown below by way of example, with an increase in the
ventilation pressure whilst the ventilation volume remains
constant.
mbar
Ventilation pressure progress before and after
the reduction in lung compliance
30
20
10
0
CPR
118).
(see "9 Alarms and error