5 Function check
Result

5.3 Preparing for the function check

Required material
Requirement
Result
52
EN
MEDUMAT Easy
Damage
Wear
Exposed wires
Bent connection lines
3. Check that all the connectors and connections engage
properly.
4. Check the testing bag for damage. Check the balloon and the
integrity of the connector.
Check that the testing bag's balloon is firmly attached to
the connector.
5. Check the patient valve, connectors, and membranes for
external damage, cracks, distortions, and soiling.
6. If necessary: Replace any damaged accessories.
7. If necessary: Dispose of any damaged accessories
Disposal", page
8.
WARNING! Device failure due to dead batteries!
Check whether a spare battery is available.
The device and accessories have been checked visually.
Oxygen supply
The device and accessories have been checked visually and are in
perfect condition
(see "5.2 Visually checking the device and
accessories", page
1. Connect the device to the oxygen supply
an oxygen supply", page
2. Connect the patient hose system and MEDUtrigger up to the
device
(see "4.6 Connecting the patient hose system and
MEDUtrigger to the device", page
3. Keep a testing bag available for subsequent steps.
The device is ready for the function check.
CPR
131).
51).
44).
(see "11.3
(see "4.4 Connecting
48).