5 Function check
Result
5.5.4 Checking the tidal volume and aware pressure
58
EN
MEDUMAT Easy
2. Count the number of inspiration phases for exactly 1 minute.
Check whether the ventilation rate is between 23 and
27 breaths per minute.
In combination with the testing bag, these settings can cause the
/
pAW
Apnea alarm to be triggered. The alarm can be ignored
during this test step.
The ventilation rate has been checked.
measurement
1. Select the following settings:
Via the adjusting knob: Vt = 950 ml at rate = 10 breaths per
minute
Via the pMax button with control LEDs: 45 mbar
In combination with the testing bag, these settings can cause the
pAW
alarm to be triggered. The alarm can be ignored during
this test step.
CPR
1:00 min
23-27
/ min