Weinmann MEDUMAT Easy CPR Instructions For Use Manual page 111

Ventilators
Table of Contents

Advertisement

9.
10. If necessary: Allow the ventilation hose to dry fully.
11. Check the pressure measuring tube for residue and any
12. If visible soiling remains: Repeat the manual cleaning.
Result
The pressure measuring tube is cleaned manually.
8.3.3 Wipe disinfecting parts
Requirement
1. The parts approved for wipe disinfection can be found in the
2. The disinfectants, doses, and exposure time for the individual
3.
4.
5. Wipe-disinfect the parts with one of the agents listed
CAUTION! Risk of injury due to disinfectant or
cleaning agent residues in the device or hose system!
Allow the pressure measuring tube to dry fully.
remaining soiling.
The parts intended for wipe disinfection have been manually
cleaned
(see "8.3.2 Cleaning parts manually", page
cleaning and disinfection plan
disinfection plan", page
parts can be found in the cleaning and disinfection plan
"8.2 Cleaning and disinfection plan", page
Warning! Risk of injury and damage to the device
caused by ingress of liquids!
Do not remove the compressed gas hose from the compressed
gas connection.
CAUTION! Risk of injury and damage to the device
caused by ingress of liquids!
Turn the connection for the ventilation hose downward.
Cleaning and disinfection plan", page
When doing so, note:
8 Cleaning and disinfection
(see "8.2 Cleaning and
100).
100).
CPR
MEDUMAT Easy
107).
(see
100).
(see "8.2
EN
111