Part
Reusable ventilation
hose with patient
valve and reusable
pressure hose
Hose protection
sleeve
Single-patient hose
with single-patient
valve and single-
patient pressure tube
Manual
Wipe
cleaning
disinfection
Immerse in
®
neodisher
MediClean
forte (Dr.
Weigert)
Dosage:
Not permitted
10 ml/l
Duration: for
at least
10 min. until
all surfaces are
visibly clean
Not permitted Not permitted
Disposable item, do not reuse. Dispose of correctly
8 Cleaning and disinfection
Machine
Immersion
cleaning and
disinfection
disinfection
Cleaning:
neodisher
MediClean
forte (Dr.
Immerse in
Weigert):
®
gigasept
FF
0.5%, 55°C,
neu (Schülke)
10 min.
Dosage:
5 ml/l
thermal
Exposure time:
disinfection:
15 min.
90°C, 5 min.
(corresponding
to A0 value
3000)
Washing at
60°C in
industrial
washing
machine with
cleaning agent:
Derval SOLO
(RKI) (Kreussler)
Dosage:
2 ml/l
Not permitted
and
Disinfectant:
Ottalin
PERACET
(Kreussler)
Dosage: 2 ml/l
Exposure time:
10 min., type
AB
(see 11.3, p.
MEDUMAT Easy
Steam
sterilization
Optionally
®
permissible:
steam
sterilization*
following
disinfection:
5 min. at
134°C
or
4 min. at
132°C
Not permitted
131).
CPR