Use a new cloth for each cleaning procedure.
All surfaces must be wiped down carefully.
All surfaces must be wetted with cleaning solution.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
Rewipe uneven surfaces and grooves in particular (e.g., top
and bottom of MEDUtrigger, adjusting knob, connection
for ventilation hose).
8. If the parts need to be immersed acc. to the cleaning and
disinfection plan: Immerse parts in the cleaning solution.
When doing so, note:
CAUTION! Risk of injury and damage to the device
as a result of unsatisfactory hygienic reliability!
The fastening strap must be detached from the mounting
plate
(see "7.7 Disconnecting the mounting plate and
fastening strap from the device", page
Open and close the fastening strap buckle repeatedly in the
immersion disinfection solution.
The fastening strap buckle must be opened.
The fastening strap around the buckle must be slackened.
Swirl the parts in the cleaning solution to coat all surfaces
and any cavities completely.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
9. If visible soiling remains: Repeat the manual cleaning.
10.
CAUTION! Risk of injury due to disinfectant or
cleaning agent residues in the device or hose system!
Rinse parts that have been immersed in the cleaning solution
thoroughly with water of drinking-water quality.
11. Wipe down the remaining parts with a damp cloth in order to
remove cleaning agent residue.
12. Wipe MEDUtrigger with a dry cloth.
8 Cleaning and disinfection
97).