The device is switched off.
1. Open the oxygen cylinder slowly.
2. Connect the testing bag to the inspiration side of the patient
valve.
3. Switch on the device using the On/Off button.
Upon being switched on, the device performs an automatic
self-test which takes approx. 2 seconds.
4. Monitor the self-test and also check the following signals:
•
Check whether the device emits an audio signal when
being switched on.
4
3
2
•
Check that all the LEDs in the control panel light up at least
once when being switched on: Demand flow (1), pMax (2),
ventilation pressure display (3), alarm (4), alarm mute (5),
Man.(6).
5 Function check
CPR
MEDUMAT Easy
5
6
1
1 *
EN
55