12. Check the pressure measuring tube for residue and any
13. If visible soiling remains: Repeat the manual cleaning and
Result
The pressure measuring tube has been immersion disinfected.
8.3.5 Machine cleaning and disinfecting parts
Alternatively to manual cleaning and disinfection, certain parts can
also be machine cleaned and disinfected.
Requirement
The cleaning and disinfection is prepared
cleaning and disinfection", page
1. The parts approved for machine cleaning and disinfection can
2. Place the parts in a washer and disinfector. When doing so,
3. Add cleaning agent and, if necessary, neutralizing agent in
4. Start machine reprocessing cycle acc. to cleaning and
5.
6. Check the parts for residue and any remaining soiling.
7. If visible soiling remains: Repeat the machine cleaning and
remaining soiling.
immersion disinfection.
be found in the cleaning and disinfection plan
Cleaning and disinfection plan", page
note:
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
Connect the hoses to the washer and disinfector using the
corresponding hose connector.
All parts and cavities must be completely flushed.
The water must be able to run away.
accordance with the washer and disinfector's instructions for
use.
disinfection plan.
CAUTION! Risk of injury due to disinfectant or
cleaning agent residues in the device or hose system!
Allow parts to dry fully at room temperature.
disinfection.
8 Cleaning and disinfection
(see "8.3.1 Preparing
106).
100).
CPR
MEDUMAT Easy
(see "8.2
EN
115