8 Cleaning and disinfection
Requirement
114
EN
MEDUMAT Easy
Immersion disinfecting the pressure measuring tube
The pressure measuring tube is cleaned manually.
1. The disinfectants, doses, and exposure time can be found in
the cleaning and disinfection plan
disinfection plan", page
2. Prepare the immersion disinfection solution in accordance with
the specifications of the disinfectant agent manufacturer.
3. Connect a sterile disposable syringe (20 ml) to a free end of the
pressure measuring tube.
4. Draw the disinfectant solution up through the pressure
measuring tube into the disposable syringe by means of suction
until both are completely full.
5. Disconnect the disposable syringe from the pressure measuring
tube.
6. Immerse the pressure measuring tube in the immersion
disinfection solution.
When doing so, note:
All surfaces and lumens must be completely wetted.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
7. Rinse the pressure measuring tube with drinking-water quality
from outside.
8.
CAUTION! Risk of injury due to disinfectant or
cleaning agent residues in the device or hose system!
Rinse the pressure measuring tube with drinking-water quality
from inside at least eight times with a disposable syringe.
When doing so, note: Only rinse in one direction.
9.
CAUTION! Risk of injury due to disinfectant or
cleaning agent residues in the device or hose system!
Allow the pressure measuring tube to dry fully.
10. If necessary: Allow the ventilation hose to dry fully.
11. If necessary: Dry hoses with sterile compressed air or medical
oxygen.
CPR
(see "8.2 Cleaning and
100).