Securing The Mounting Plate With Fastening Strap To The Device - Weinmann MEDUMAT Easy CPR Instructions For Use Manual

Ventilators
Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

1. Install the device on the carrying system or carrying structures
using screws and protective disks.
2. Tighten screws.
Result
The device is installed on a carrying system or carrying structures.
4.3 Securing the mounting plate with
fastening strap to the device
Malfunction or failure of treatment as a result of insufficiently
secured devices!
If the device is not sufficiently secured, its uncontrolled movement
can lead to functional failure. This can result in serious or life-
threatening injury to the patient.
 Always pull the fastening strap tight.
CPR
MEDUMAT Easy
4 Preparation