6.4.2 Ventilating the patient in manual mode
(CPR mode)
Lack of treatment due to the attempt to trigger manual
breaths in demand mode and continuous mode!
If the device is not in manual mode (CPR mode) and the
MEDUtrigger is not connected, the device/user cannot manually
trigger mechanical breaths. This can injure the patient.
 For manual mechanical breaths, connect MEDUtrigger and
Switching on manual mode
Requirement
switch to manual mode (CPR mode).
The device is ready for use
page
64).
MEDUtrigger is connected
hose system and MEDUtrigger to the device", page
A tidal volume is set with the adjusting knob for the ventilation
parameters
(see "6.2.1 Setting the respiratory rate and tidal
volume", page
70).
The device is switched on
page
72).
Voice prompt is deactivated
the voice prompt", page
The metronome is deactivated
deactivating the metronome", page
(see "6.1 Preparing for ventilation",
(see "4.6 Connecting the patient
(see "6.3 Switching the device on",
(see "6.7 Activating/deactivating
85).
(see "6.8 Activating/
87).
CPR
MEDUMAT Easy
6 Operation
48).
EN
75