6 Operation
Result
Result
80
EN
MEDUMAT Easy
A new mechanical breath can only be triggered if the inspiration
and expiration phase of the previous mechanical breath has
already ended.
An information tone sounds if the inspiration and expiration
phase has not yet ended.
If the button on the MEDUtrigger is pressed continuously, the
device triggers an unlimited number of subsequent mechanical
breaths.
4. Perform 30 chest compressions and two mechanical breaths in
alternation.
The voice prompts and metronome are static. We recommend
adjusting to the metronome and voice prompts.
The device is operated in CPR mode.
Switching off CPR mode
1. Press the Man. button.
The device continues to ventilate in continuous mode.
or
2. Activate the demand flow mode
patient in demand flow mode", page
or
3. Switch off the device
page
84).
The CPR mode is switched off.
CPR
(see "6.4.1 Ventilating the
73).
(see "6.6 Switching the device off",