9 Alarms and error messages
Alarm
/
Alarm pAW
Apnea (respiratory
system interrupted)
120
EN
MEDUMAT Easy
Possible cause
Patient hose leaking/
slipped off/kinked
Ventilation mask/tube
not seated correctly/
kinked
Pressure measuring
tube leaking/slipped
off/kinked
The ventilation mask is
connected to the
expiration side of the
patient valve
The patient has no/
insufficient
spontaneous
respiration in demand
flow mode
Defective MEDUtrigger
/ MEDUtrigger button
continually pressed in
manual mode
MEDUtrigger is not
actuated in manual
mode
Device defective
CPR
Remedy
Check connections and the patient
hose system and remedy any leaks.
Attach the ventilation mask to the
inspiration side
(see 5.5, p.
Switch to manual mode (CPR mode) or
continuous mode
(see 6.4, p.
Release the MEDUtrigger button
Check MEDUtrigger
If possible, switch to continuous
mode (see 6.4.3, p. 72)
Replace the device
Perform ventilation
If necessary: Switch to continuous
mode
Have the device repaired
p.
6).
To be
remedied by
User
User
54).
User
73).
User
User
Manufacturer or
(see 1.2,
authorized
technician