Part
Device
Mounting plate
Compressed gas tube
MEDUtrigger
Manual
Wipe
cleaning
disinfection
Wipe-down
cleaning with
®
neodisher
MediClean
forte (Dr.
Weigert)
Dosage:
Wipe down
10 ml/l,
with
Duration: treat
Incidin
all surfaces at
OxyWipes S
least twice
(Ecolab)
until visibly
Exposure time:
clean
5 min.
or
or
®
CIDEZYME
Super Sani-
®
Enzymatic
Cloth
Detergent
Disposable
Solution
Wipes (PDI
(Johnson&
Healthcare)
Johnson)
Exposure time:
Dosage:
2 min.
16 ml/l
Duration: for
at least 3 min.
until all
surfaces are
visibly clean
Immersion
disinfection
TM
Not permitted
MEDUMAT Easy
8 Cleaning and disinfection
Machine
Steam
cleaning and
sterilization
disinfection
Not permitted
Not permitted