Weinmann MEDUMAT Easy CPR Instructions For Use Manual page 71

Ventilators
Table of Contents

Advertisement

6.2.2 Setting the maximum ventilation pressure
Requirement
The device is switched on
page
1. Set the ventilation pressure. To do this, press the pMax button
Result
The maximum ventilation pressure is set.
Recommendation for maximum ventilation pressure
Mask ventilation
20 mbar (20 cmH
If, for example, with reduced lung compliance the set maximum
ventilation pressure is reached, the device emits the pAW
72).
with control LEDs.
The associated LED displays the set maximum ventilation
pressure.
O)
2
(see "6.3 Switching the device on",
Tube ventilation
45 mbar (45 cmH
MEDUMAT Easy
6 Operation
O)
2
alarm.
CPR
EN
71