3. Disassembling the reusable patient valve:
•
Remove the spontaneous respiration insert (1) from the
spontaneous respiration side (3) of the reusable patient
valve.
To do so, push both locking tabs out of the mount using a
small slot-head screwdriver.
•
Remove the O-ring (9) from the spontaneous respiration
insert (1).
•
Remove the disk diaphragm/emergency air membrane (2)
from the spontaneous respiration side (3) of the reusable
patient valve with pointed pliers.
•
Remove the disk diaphragm (4) from the expiration side (5)
of the reusable patient valve with pointed pliers.
•
Unscrew the connection for ventilation hose (6) from the
reusable patient valve lid (7) in a counterclockwise
direction.
•
Unscrew the reusable patient valve lid (7) in a
counterclockwise direction.
3
MEDUMAT Easy
7 Disassembly