3. Disassembling the reusable patient valve:
Remove the spontaneous respiration insert (1) from the
spontaneous respiration side (3) of the reusable patient
valve.
To do so, push both locking tabs out of the mount using a
small slot-head screwdriver.
Remove the O-ring (9) from the spontaneous respiration
insert (1).
Remove the disk diaphragm/emergency air membrane (2)
from the spontaneous respiration side (3) of the reusable
patient valve with pointed pliers.
Remove the disk diaphragm (4) from the expiration side (5)
of the reusable patient valve with pointed pliers.
Unscrew the connection for ventilation hose (6) from the
reusable patient valve lid (7) in a counterclockwise
direction.
Unscrew the reusable patient valve lid (7) in a
counterclockwise direction.
3
MEDUMAT Easy
7 Disassembly
7
8
9
1
2
CPR
EN
95