7 Disassembly
Result

7.6 Removing the oxygen supply

Requirement
96
EN
MEDUMAT Easy
Remove the lip membrane (8).
The reusable hose system is disassembled.
The device is switched off
page
84).
The oxygen supply is closed.
1. NOTICE! Material damage due to the disconnection of a
non-ventilated system!
Switch on the device using the On/Off button.
The remaining oxygen can escape and the device is
depressurized. The compressed gas hose can only be
disconnected without tools if the contents gauge indicates
0 bar (0 psi) on the pressure reducer.
2. Switch off the device using the On/Off button. To do this, keep
the On/Off button depressed until all 4 alarm LEDs light up.
Then release the On/Off button.
3
2
1
3. NOTICE! Material damage due to the use of a tool!
Disconnect the compressed gas tube (1) from the pressure
reducer (3) or the central gas supply by hand using the union
nut (2) or the quick-release coupling. To do this, turn the
knurled union nut (2) counterclockwise.