7 Disassembly
Result
7.6 Removing the oxygen supply
Requirement
96
EN
MEDUMAT Easy
•
Remove the lip membrane (8).
The reusable hose system is disassembled.
•
The device is switched off
page
84).
•
The oxygen supply is closed.
1. NOTICE! Material damage due to the disconnection of a
non-ventilated system!
Switch on the device using the On/Off button.
The remaining oxygen can escape and the device is
depressurized. The compressed gas hose can only be
disconnected without tools if the contents gauge indicates
0 bar (0 psi) on the pressure reducer.
2. Switch off the device using the On/Off button. To do this, keep
the On/Off button depressed until all 4 alarm LEDs light up.
Then release the On/Off button.
3
2
1
3. NOTICE! Material damage due to the use of a tool!
Disconnect the compressed gas tube (1) from the pressure
reducer (3) or the central gas supply by hand using the union
nut (2) or the quick-release coupling. To do this, turn the
knurled union nut (2) counterclockwise.