1. Press the Man. button.
The LEDs on the Man. button and on the MEDUtrigger
display the operational status.
The continuous ventilation with the set rate stops.
The device outputs the message: CPR mode switched on,
manual triggering of ventilation.
If the metronome is switched on, the device outputs the message:
Perform chest compressions now!
2. Perform 30 chest compressions.
When doing so, note: The metronome sets the ideal frequency.
The tone pitch increases with the metronome's last three
strikes.
The device outputs the message:
Provide 2 ventilations now!
3. Trigger two mechanical breaths:
Press and hold the button on the MEDUtrigger until two
mechanical breaths have been triggered.
or
Press the button on the MEDUtrigger twice in a row.
CPR
MEDUMAT Easy
6 Operation