1. Press the Man. button.
•
The LEDs on the Man. button and on the MEDUtrigger
display the operational status.
•
The continuous ventilation with the set rate stops.
•
The device outputs the message: CPR mode switched on,
manual triggering of ventilation.
If the metronome is switched on, the device outputs the message:
Perform chest compressions now!
2. Perform 30 chest compressions.
When doing so, note: The metronome sets the ideal frequency.
The tone pitch increases with the metronome's last three
strikes.
The device outputs the message:
Provide 2 ventilations now!
3. Trigger two mechanical breaths:
Press and hold the button on the MEDUtrigger until two
mechanical breaths have been triggered.
or
Press the button on the MEDUtrigger twice in a row.
CPR
MEDUMAT Easy
6 Operation
EN
79