5 Function check
Result
5.5.6 Checking the demand flow mode
62
EN
MEDUMAT Easy
4. Exit manual mode (CPR mode). To do this, press the Man.
button again.
The MEDUtrigger has been checked.
1. Select the "demand flow" setting. To do this, turn the
adjusting knob for ventilation parameters clockwise past the
snap-in position (1).
•
Check whether the green demand flow LED (2) lights up.
CPR
1
2