6 Operation
Result
88
EN
MEDUMAT Easy
3. If the 45 mbar or 50 mbar LED lights up on the ventilation
pressure display: Release the Man. button.
The device displays the metronome's operating status.
•
50 mbar LED lights up red: Metronome is deactivated.
•
45 mbar LED lights up green: Metronome is activated.
4. To change the metronome's operating status: Press the Man.
button.
5. To confirm the metronome's operating status: Press the On/Off
button.
A confirmation sounds:
•
If the confirmation sounds once: Deactivated metronome is
confirmed.
•
If the confirmation sounds twice: Activated metronome is
confirmed.
The metronome is activated or deactivated.
CPR