13 Technical data
Specification
Application parts
Essential characteristics
Electromagnetic compatibility (EMC) as
per EN 60601-1-2:
Radio interference suppression
Radio interference immunity
Resistance to shock and vibration
Type of rescue vehicle
The device satisfies all the relevant parts
of the following standards
UL certification
136
EN
MEDUMAT Easy
Device
Ventilation mask
Ventilation tube
Delivery of ventilation volume or triggering of alarm status
Limitation of maximum ventilation pressure
Test parameters and limit values are available on request from
the manufacturer (WEINMANN Emergency Medical Technology
GmbH + Co. KG, Frohbösestraße 12, 22525 Hamburg,
GERMANY).
EN 55011, Group 1, Class B
UN/ECE R 10, Annexes 6 and 7
RTCA DO 160 G, Section 21, Category M
EN 61000-4-2, contact ± 8kV, air ± 15 kV
EN 61000-4-3, 30 V/m to 2.7 GHz (ISM bands up to
approx. 6 GHz)
EN 61000-4-6, 3 Vrms (ISM bands with 6 Vrms)
EN 61000-4-8, 30 A/m
EN 1789
EN 60601 60601-1-12 (Categories: Secured in a rescue
vehicle, secured in an aeroplane, secured in a helicopter,
portable at the site of the emergency)
EUROCAE ED-14G (RTCA DO 160 G) (Section 8 - Vibration
Aeroplanes Cat. S and Helicopter Cat. U2)
Secured in rescue vehicle, aeroplane and helicopter as well as
portable at the site of the emergency
DIN EN 794-3
ISO 10651-3
EN 1789
EN 60601-1
EN 60601-1-2
EN 60601-1-12
RTCA DO-160 G
Regarding protection against electric shock, fire prevention, and
protection from mechanical hazards
ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012)
CAN/CSA-C22.2 No, 60601-1 (2014)
IEC 60601-1-6:2010, AMD1:2013
IEC 60601-1-8:2006, AMD1:2012
IEC 60601-1-12:2014
(1)
STPD = Standard Temperature and Pressure, dry
(21°C, 1.013 mbar)