Safe Use Of The Device And Accessories - Weinmann MEDUMAT Easy CPR Instructions For Use Manual

Ventilators
Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

2 Safety
2.4.2 Disposable items

2.5 Safe use of the device and accessories

2.5.1 Preventing interference between the devices
12
EN
MEDUMAT Easy
disinfection plan
page
100).
 Wear suitable protective equipment (e.g., gloves) during
cleaning and disinfection.
 Only use the specified products for cleaning and disinfection.
Re-using disposable items causes the following risks:
Infection if items come into contact with airways
Malfunctions when using the device
Unforeseeable reactions as a result of aging, embrittlement,
wear, thermal load and the effects of chemical processes
 Never perform cleaning and disinfection of disposable items.
 Never use disposable items more than once.
Electrical devices which are operated directly next to or on top of
each other can cause mutual interference to functionality. Portable
high-frequency communication devices in the direct vicinity of the
device can also influence the functioning of the device.
 Do not stack the device with other electrical devices.
 Do not operate the device directly next to other electrical
devices. Exception: Other WEINMANN Emergency devices
which have been tested and shown to guarantee interference-
free operation with the adjacent device. A list of other devices
is available on request.
 If stacking or operation in the immediate vicinity cannot be
avoided: Closely monitor the functioning of all affected
medical electrical devices and do not use if functions are
disrupted.
 With portable RF communication devices, maintain a minimum
distance of 30 cm (approx 12 inches) to the device and
accessories. Examples: Wireless device, mobile telephone.
CPR
(see "8.2 Cleaning and disinfection plan",