Risk of injury and material damage due to disinfectant or
cleaning agent residues in the device or hose system!
Disinfectant or cleaning agent residues can get into the patient's
lung. This can injure the patient and cause material damage to the
device.
 Never immerse the device in disinfectant or other liquids.
 Only use accessories approved for immersion disinfection.
 Turn the connection for the ventilation hose downwards during
cleaning and disinfection so that no liquids can get into the
device.
 Keep the battery compartment closed during cleaning so that
no liquids can get into the device.
 Do not remove the compressed gas tube from the compressed
gas connection so that no liquids can get into the device.
 All parts of the hose system must be rinsed thoroughly with
water and allowed to dry completely following the cleaning
and disinfection.
 Following cleaning and disinfection, visually inspect the device
and hose system for any cleaning agent residue and remove
any residue if necessary.
 Perform a full function check after every cleaning and
disinfection.
Material damage due to the ingress of liquids!
The device is protected against the ingress of water. This only
applies when the battery compartment is closed. The ingress of
liquids may cause damage to the device and accessories.
 Do not immerse the device in liquids.
 Only use accessories approved for immersion disinfection.
 Only clean the battery compartment with a dry cloth to avoid
liquids entering the device.
8 Cleaning and disinfection
CPR
MEDUMAT Easy
EN
105