8 Cleaning and disinfection
Result
8.3.4 Immersion disinfecting parts
Requirement
112
EN
MEDUMAT Easy
Wet any uneven surfaces and grooves (e.g., adjusting
knob, connection for ventilation hose) sufficiently with the
disinfectant.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
6.
CAUTION! Risk of injury due to disinfectant or
cleaning agent residues in the device or hose system!
Allow the parts to dry fully.
7. Check the parts for residue and any remaining soiling.
8. If visible soiling remains: Repeat the cleaning and wipe
disinfection procedure.
Depending on the disinfectant, rewiping with a neutralizing agent
may be necessary.
The parts are wipe disinfected.
The parts intended for immersion disinfection have been
manually cleaned
page
107).
1. The parts approved for immersion disinfection can be found in
the cleaning and disinfection plan
disinfection plan", page
CPR
(see "8.3.2 Cleaning parts manually",
(see "8.2 Cleaning and
100).