Blank Value Correction (Bw Corr) - wtw inoLab Level 3 Operating Manual

Measuring system with active multifunction boxes
Table of Contents

Advertisement

Ion concentration
96
12
If mg/kg, ppm, % was selected:
Enter the density of the sample with
Continue with step 13.
13
Confirm the entry with
The units change over.
14
Switch to the measured value display.
7.2.2

Blank value correction (Bw Corr)

If the ion concentration in the sample is so small that it does
not lie in the linear range of the electrode, you can increase
the ion concentration in the linear range with the blank value
correction by the addition of a blank value solution.
The displayed measured value is the difference between the
actual measured value and the determined blank value.
1
Switch to the Menu mode.
46
2
Use
3
Select the ISE menu item with
4
Confirm the ISE menu item with
5
Select the ISE setup menu item with
6
Confirm the ISE setup menu item with
7
Select the Bw Corr menu item with
8
Confirm the Bw Corr menu item with
The Bw Corr function is active (✔).
The flashing
value display.
9
Perform a calibration (see section 7.3).
m
10
Use
to finish the calibration.
The prompt for the measurement of the blank value
appears.
g
or [ESC].
to select the main Meas menu.
-
symbol appears on the measured
82
.
82
.
g
.
82
.
g
.
82
.
g
.