wtw inoLab Level 3 Operating Manual page 80

Measuring system with active multifunction boxes
Table of Contents

Advertisement

pH
Calibration interval
80
The calibration interval determines the time interval between
regular system calibrations.
The terminal reminds you to calibrate regularly by the
flashing probe symbol on the display.
The calibration interval is set to 7 days in the factory. You
can change the interval (1-999 days).
For most applications, we recommend a shorter calibration
interval.
1
Switch to the Menu mode.
46
2
Use
82
3
Use
to select the Cal Int. ... menu item.
g
4
Use
to confirm the Cal Int. ... menu item.
The Calibration interval entry window appears.
5
Option:
Switch between the individual digits with
6
Increase or reduce the value with
7
Confirm the value for the calibration interval with
8
Use [ESC] to terminate the Menu mode.
to select the main Cal menu.
46
.
82
.
g
.