® SenTix Technical data Technical data General data WTW model Reference electrolyte Junction NTC Special features ® SenTix KCl 3 mol/L, Ag free Ceramic Plastic shaft ® SenTix KCl 3 mol/L, Ag free Ceramic Plastic shaft ® SenTix KCl 3 mol/L, Ag...
Page 5
® SenTix Commissioning, measuring, calibration Commissioning, measuring, calibration Commissioning The combination electrode is filled with reference electrolyte solution in the factory. Prepare the combination electrode for measuring as follows: Open the refilling opening for the reference electrolyte solution. Depend- ing on the model, the stopper of the refilling opening is an elastomer stop- per or a slider.
Page 6
® Storage SenTix of the shaft (see figure). At the same time, the level of the reference elec- trolyte must be at least 2 cm above the level of the solution. Reference electrolyte Shaft Junction (Example) Glass membrane Provide approximately the same stirring conditions for measuring as for calibrating.
® SenTix Aging Aging Every pH combination electrode undergoes a natural aging process. With ag- ing, the responding behavior becomes slower and the electrode slope and asymmetry change. Moreover, extreme operating conditions can consider- ably shorten the lifetime of the combination electrode. These are: ...
108 112 Connection cable S7 plug-in connector/BNC, AS/BNC 108 114 Note Detailed information on our wide range of buffer solutions and more acces- sories is given in the price list of the WTW catalog "Laboratory and field in- strumentation". ba75737e02 06/2013...
Page 12
For more information on how Xylem can help you, go to www.xyleminc.com Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: +49 881 183-0 Fax: +49 881 183-420 E-Mail: info@wtw.com Internet: www.wtw.com...