wtw inoLab Level 3 Operating Manual page 128

Measuring system with active multifunction boxes
Table of Contents

Advertisement

Ion concentration
Measuring
128
1
Select the measuring method (see section 7.4.1).
2
Prepare two standard calibration solutions.
3
Perform a two-point calibration according to user
guidance.
4
As soon as a stable value for the second standard
calibration solution is reached, the
display indicator
the electrode is displayed after the two-point
calibration. The probe symbol shows the electrode
evaluation after the two-point calibration.
5
Start the measurement with [Start].
The entry for Standard volume, Standard
concentration and ISA volume opens.
6
Enter Standard volume, Standard concentration and
ISA volume with
7
Thoroughly rinse the electrode with deionized water.
8
Immerse the electrode in the standard solution.
h
stops flashing. The slope (mV) of
82
and [Next].