Two-Point And Three-Point Calibration - wtw inoLab Level 3 Operating Manual

Measuring system with active multifunction boxes
Table of Contents

Advertisement

7.3.1
Two-point
calibration

Two-point and three-point calibration

1
Perform the preparatory activities.
2
Switch to the ISE display with
c
3
Press the
of the first standard solution is displayed.
82
4
Use
to enter the concentration of the first
standard solution.
5
Immerse the electrode in standard solution 1.
g
6
Press the
The electrode voltage is displayed and the
display indicator flashes.
7
As soon as a stable value is reached,
flashing and the field to input the concentration of the
second standard solution appears.
m
key. A field to select the concentration
key.
Ion concentration
46
and/or
.
h
h
stops
105