BOD
158
38
Enter the total volume of the sample dilution in the
Vol.Total field with
39
Immerse the sensor in the sample dilution.
40
Start the measurement with [Start].
The sample dilution is measured.
41
Wait for the end of the measurement.
42
Repeat from step 33 to step 42 for all sample
dilutions.
Subsequently, the Routine STORE window opens.
43
Select Save Routine YES or No in the field with
If Save Routine YES was selected, the routine is
saved under a routine name.
44
Confirm the settings with [Ok].
l
If Save Routine NO was selected, the start
measurement is ended.
l
If Save Routine YES was selected, an entry field
for the routine name appears.
82
or an external keyboard.