What to do if...
Test procedure
Control unit is switched
- no display indication
- no reaction
Display of the control
unit works. Desired
temperature is not
achieved. The actual
temperature value is too
18
6.2

Measuring the actual value of the sample temperature

1
Place a BOD sample bottle filled with approx. 400 ml water on
a grating in the center of the thermostat cabinet.
2
Completely immerse the probe of a temperature meter (accu-
racy at least 0.5 °C, resolution at least 0.1 °C) in the sample
bottle. The probe must not touch the bottle wall.
3
After three hours adjustment time read the temperature value.
6.3

Error diagnosis

Cause
on,
Mains power supply failed
Fuse of the control unit defective
Control unit defective
Cause
The waiting time was not sufficient
for the loading of the thermostat
cabinet.
low
The ambient temperature is too
low.
Control unit defective, compres-
sor runs permanently.
TS 606...-i/1006-i
Remedy
Check mains power supply, con-
nect an operable consumer (e.g.
lamp) to the power socket
Check the fuse of the control
unit. If it is defective, replace it by
a new fuse (value, see chapter 7
T
).
ECHNICAL DATA
The fuse is available in specialist
shops.
Warning: Only use the specified
type. Do not repair fuses!
Contact the WTW service de-
partment.
Remedy
Wait for a longer period of time
and observe the development of
the actual value.
Check the ambient temperature.
Contact the WTW service de-
partment.