wtw inoLab Level 3 Operating Manual page 153

Measuring system with active multifunction boxes
Table of Contents

Advertisement

Start measurement
without routine
1
Switch to the Menu mode.
46
2
Use
to select the main Meas menu.
3
Select the BOD menu item with
4
Confirm the BOD menu item with
5
Select the Start Meas menu item with
6
Confirm the Start Meas menu item with
The Start measurement window appears.
l
To start a measurement without routine, continue
at step 7.
l
To start a measurement with routine, continue at
step 15.
7
Select No in the With routine field using
The entry field New dilution water appears.
8
Swap between the entry fields with [Next].
9
Select Yes or No in the New dilution water field with
82
.
l
If New dilution water No was selected: The
Number entry field disappears. The measurement
of the blank value solutions is skipped. The blank
value solutions last measured are used for the
BOD determination.
l
If New dilution water Yes was selected: The
Number entry field appears. A new blank value
solution is measured to determine the BOD value.
BOD
82
.
g
.
82
.
g
.
82
.
153