7.4.2

Standard Addition (Std Add)

In the "Standard addition" procedure, a known amount of
standard solution is added to the sample.
The ion concentration in the sample is calculated from the
change in potential.
1
Select the measuring method (see section 7.4.1).
2
Prepare two standard calibration solutions.
3
Perform a two-point calibration according to the user
guidance.
4
As soon as a stable value for the second standard
calibration solution is reached, the
display indicator
the electrode is displayed after the two-point
calibration. The probe symbol shows the electrode
evaluation after the two-point calibration.