Ion concentration
116
10
Select the ISE Adv. menu item with
11
Confirm the ISE Adv. menu item with
The menu with the measuring methods opens.
12
Select a method with
13
Confirm the method with
Measurement with the selected method begins.
7.4.2
Standard Addition (Std Add)
In the "Standard addition" procedure, a known amount of
standard solution is added to the sample.
The ion concentration in the sample is calculated from the
change in potential.
1
Select the measuring method (see section 7.4.1).
2
Prepare two standard calibration solutions.
3
Perform a two-point calibration according to the user
guidance.
4
As soon as a stable value for the second standard
calibration solution is reached, the
display indicator
the electrode is displayed after the two-point
calibration. The probe symbol shows the electrode
evaluation after the two-point calibration.