Data storage
196
11.1.2 Automatically storing measurement data
records
AutoStore enables measurement data records to be stored
at defined time intervals. Starting AutoStore causes you to
move to the Set AutoStore input window where you define
the details of the storage with AutoStore.
l
ID:
The name (ID) identifies the measurement and makes it
easier to find the data in memory. The ID can consist of
figures and letters (when using an external IBM PC
keyboard).
l
Duration:
Here you set up the measuring period for which the
planned AutoStore storing is to be performed.
l
Interval:
Here you determine the intervals at which the measuring
data records will be stored.
l
Memory requirements (datasets):
Here you obtain information on how many storage
locations are required with the current settings.
l
Available memory locations:
Here you obtain information on how many storage
locations are still available.
Note
If insufficient memory locations are available, a message
appears on the display.
Change the entries in the Duration and/or Interval field until
an available number of memory locations is displayed.