Preparatory
activities
Perform the following preparatory activities:
1
Switch on the measuring system with
2
Connect the electrode to the measuring module.
3
Connect a temperature probe to the measuring
system so that any temperature differences that are
too high will be indicated during calibration.
4
Keep the standard solutions ready.
5
Adjust the temperature of the standard solutions.
6
Switch to the Menu mode.
46
7
Use
8
Confirm the ISE menu item with
9
Confirm the ISE menu item with
The measured variable ISE appears on the display
(✔).
Note
For calibration, use tempered standard solutions with a
temperature that differs by a max. of
temperature of the test sample.
to select the main Meas menu.
Ion concentration
e
.
82
.
g
.
2 °C from the
±
103