BOD
162
sample dilutions is acceptable. Reading in the identification
automatically causes the correct assignment of the start
measurement to the measured value.
In an end measurement without a barcode reader, follow the
prescribed order of the measurements.
12
Check the data on the display.
l
If the data is correct: Continue with step 16.
l
To change the data: Continue with step 13.
13
Swap between the selection fields with [Next].
14
Select Yes or No in the Remove field with
l
If Remove No was selected, the (default) values
from the start measurement are adopted.
l
If Remove Yes was selected, no end
measurement is performed for this blank value
solution. The value appears on the report marked
by *. This value is not used for calculating the BOD
value.
15
Immerse the sensor in the blank value solution.
16
Confirm the settings with [Start].
l
If Remove Yes was selected, a security prompt
appears before the removal. After confirmation
with Yes, the next blank value solution is offered
for the end measurement (see step 12).
l
If Remove No was selected, the measurement of
the blank value solution begins.
82
.