wtw inoLab Level 3 Operating Manual page 31

Measuring system with active multifunction boxes
Table of Contents

Advertisement

8
Plug the power supply into the inoLab Level 3
terminal again.
The Standby display appears.
g
9
Press
and hold it down.
10
Press [Offline].
11
Release both keys.
The Terminal Programming window shows the
current status of the update (in %).
12
Wait for display to reach 100%.
The firmware update has been carried out.
13
Briefly unplug the inoLab Level 3 terminal from the
supply voltage and plug it again.
The Standby display appears. The instrument is
ready for operation.
Commissioning
31