Operating instructions
Water sampler
PB 150/2 & PB150-SE
ba75817e01
05/2009

Advertisement

Table of Contents
loading

Summary of Contents for wtw PB 150/2

  • Page 1 Operating instructions Water sampler PB 150/2 & PB150-SE ba75817e01 05/2009...
  • Page 2 MAXX GmbH. Any reproduction, use or duplication of this manual, also in extracts, is prohibited by law without the previous written agreement of MAXX GmbH. Subject to changes! Access code for programming, system settings, key lock Password: 6299 Your password: ba75817e01 05/2009 - 2 - PB 150/2 & PB 150-SE...
  • Page 3: Table Of Contents

    MESSAGES............................73 DIGITAL INPUTS / OUTPUTS ......................74 Spare parts / accessories PB 150/2 ....................88 Circuit diagram mains connection PB 150/2 & PB150-SE ..............89 Circuit diagram vacuum system......................90 Circuit diagram bypass system......................91 Circuit diagram flow-proportional system................... 92 Circuit diagram terminal connections....................
  • Page 4 General Information Circuit diagram PB 150/2-A 12 x 1,6 L flow-proportional ..............98 Circuit diagram PB 150/2-A 12 x 1,6 L flow-proportional ..............99 Circuit diagram PB 150-SE 2 x 10 L vacuum ................100 Circuit diagram optional inlet / additional valve 900276 ..............101 Circuit diagram optional inlet / additional valve 900276 ..............
  • Page 5: General Information

    The sampler can be operated at ambient temperatures from -20° C to +43°C. Sampling from pressurised lines is not possible without optional accessories. The device is weather-proof and suitable for outside operation. ba75817e01 05/2009 - 5 - PB 150/2 & PB 150-SE...
  • Page 6: Redelivery For Repair

    All packing materials can be disposed of as usual. Packing materials are: cardboard, wood, PS and PE. If the packing is returned free of cost, we will take care of its disposal. ba75817e01 05/2009 - 6 - PB 150/2 & PB 150-SE...
  • Page 7: Use Of Hazard Information

    Producer for disposal at no charge to the user. Note: For all electrical products (marked or unmarked) which are supplied or produced by MAXX, please contact the local MAXX sales office for instructions for proper disposal. ba75817e01 05/2009 - 7 - PB 150/2 & PB 150-SE...
  • Page 8 1400 (2015*) x 930 x 850 mm for bottle variants: 4x20 L, 4x25 L, 24x2,9 L, 24x2 L, 36x1 L *) with opened top Weight 100 kg with composite container; higher weight when using several ba75817e01 05/2009 - 8 - PB 150/2 & PB 150-SE...
  • Page 9 Wetted materials PVC, Silicone, PS, PE, EPDM (optional: glass (Duran50) metering vessel, sinker weight SS304) Device designation PB 150/2-M (with measurin rack) (like PB 150/2, with the following modifications) Device type Fixed site sampler with measuring rack for measuring transducers...
  • Page 10 General Information Device designation PB 150-SE (like PB 150/2, with the following modifications) Device type Self-emptying fixed site sampler for continuous operation without the need of personnel (monitoring applications) Emptying/rinsing Bottles are emptied, rinsed and re-filled on the same bottle position.
  • Page 11: Characteristics: Sampler Pb 150/2

    The illustrated sampler corresponds to a standard version with 12 bottles and distributor plate. According to the sampler version, some components such as dosing unit or distributor/bottles can be different! ba75817e01 05/2009 - 11 - PB 150/2 & PB 150-SE...
  • Page 12: Haracteristics: Sampler Pb 150-Se

    Swivelling sample discharge tube Remark: The illustrated sampler corresponds to a standard version with 24 bottles. According to the sampler version, some components such as dosing unit or distributor/bottles can be different! ba75817e01 05/2009 - 12 - PB 150/2 & PB 150-SE...
  • Page 13: Dimensions Of Device Pb 150-Se Depending On The Bottle Variant

    General Information Dimensions of device PB 150-SE depending on the bottle variant (see technical specifications) ba75817e01 05/2009 - 13 - PB 150/2 & PB 150-SE...
  • Page 14: Installing The Apparatus

    Messages The relays required for messages (up to 4) can be fixed on the top hat rail. For details please see the circuit diagram ba75817e01 05/2009 - 14 - PB 150/2 & PB 150-SE...
  • Page 15 To replace the fuse, pull out fuse holder, open the cover and remove the fuse. 2. Fuse protection of CPU board by an automatic fuse. The fuse is reactivated by disconnecting the device from the mains. ba75817e01 05/2009 - 15 - PB 150/2 & PB 150-SE...
  • Page 16: Troubleshooting

    - Fuse defective -Check all fuses and replace them if necessary Cooling system does - Power supply -Check mains connection of not function cooling machine - Cooling system leaking -Contact service department ba75817e01 05/2009 - 16 - PB 150/2 & PB 150-SE...
  • Page 17: Connections

    1 = Rx 2 = free 3 = Tx 4 = free 5 = Gnd Terminal assignment – input signals Remark: Terminals K2 to K4 are optional depending on the device version. ba75817e01 05/2009 - 17 - PB 150/2 & PB 150-SE...
  • Page 18 In the future software version V3_51 the numbering of the inputs and outputs of the control unit when programming and in the test display will be identical with the terminal numbers on the adapter board. ba75817e01 05/2009 - 18 - PB 150/2 & PB 150-SE...
  • Page 19: Sampling Modes

    A sample is extracted in fixed time intervals (e.g. every 10 min.) with a fixed Event- volume (e.g. 50 ml) as long as an event is present. proportional ba75817e01 05/2009 - 19 - PB 150/2 & PB 150-SE...
  • Page 20 = big sample volume proportionally scaling the mA signal low flow = small sample volume. and the maximum sample volume. That means, the sample volume is automatically adjusted to the mA signal. ba75817e01 05/2009 - 20 - PB 150/2 & PB 150-SE...
  • Page 21 If an event is longer than the programmed bottle filling time, two or more bottles will be filled for this event depending on the programmed bottle filling time. ba75817e01 05/2009 - 21 - PB 150/2 & PB 150-SE...
  • Page 22 If you need to work with different sample volumes, you can cut to size several tube pieces. As the volume control tube (5) is only slipped on the hose connector (1), it is very easy to change the tube. ba75817e01 05/2009 - 22 - PB 150/2 & PB 150-SE...
  • Page 23 Wrong operating conditions can lead to malfunctions or may even damage the apparatus. No liability or claims under guarantee will be accepted in respect of any damage arising from the non-observance of these operating instructions. ba75817e01 05/2009 - 23 - PB 150/2 & PB 150-SE...
  • Page 24: Troubleshooting - Vacuum Sampling System

    Valve system/pinch valve does Check valve system/pinch valve not function correctly or contact service department No pressure / vacuum at Defective diaphragm Renew diaphragm pump Hoses are squeezed Check hoses ba75817e01 05/2009 - 24 - PB 150/2 & PB 150-SE...
  • Page 25: Flow-Proportional Sampling System

    However, a sample would have been very important for an analysis and the taking of countermeasures. Flow-proportional sampling system - structure Electrodes Air inlet Screw cap Metering vessel Sample inlet Motor-driven pinch valve Silicone discharge tube ba75817e01 05/2009 - 25 - PB 150/2 & PB 150-SE...
  • Page 26 Remark: Devices with a flow-proportional sampling system have to be operated with a suction hose ID of 16 mm. ba75817e01 05/2009 - 26 - PB 150/2 & PB 150-SE...
  • Page 27: Troubleshooting - Flow-Proportional Sampling System

    Pump does not produce Diaphragm defective Replace diaphragm pressure/vacuum Hoses are kinked Check hoses Inaccurate sample volume System has been initialized or Calibrate hardware has been changed ba75817e01 05/2009 - 27 - PB 150/2 & PB 150-SE...
  • Page 28: Bypass Sampling System

    4. Pinch valve 2 is opened. Pinch valve 1 is opened Water drains off up to the and the sample is filled level of the dosing tube. into the bottle. ba75817e01 05/2009 - 28 - PB 150/2 & PB 150-SE...
  • Page 29: Troubleshooting - Bypass Sampling System

    Metering vessel -Outlet hose is clogged or -Remove clogging overflowing (backflow) squeezed -Pinch-valve does not open -Adjust position of pinch valve completely ba75817e01 05/2009 - 29 - PB 150/2 & PB 150-SE...
  • Page 30: Sample Distribution At Pb 150/2-F-Ms

    Operation Sample distribution at PB 150/2-F-MS Rotating distributor Motor-driven rotating distributor with encoder (pulse generator) which ensures: an exact sample discharge position as well as an automatic bottle position identification by means of a forked light barrier. Remark: For reasons of operating safety, the distributor first always runs over position 1 again when changing to the next bottle.
  • Page 31: Troubleshooting - Distributor Pb 150/2-F-Ms

    Operation Troubleshooting – distributor PB 150/2-F-MS If a fault develops in your sampler, we can only provide prompt assistance if you give us the type and serial number of the apparatus concerned. You will find these data on the type plate.
  • Page 32: Sample Distribution At Pb 150-Se - Variants: 12 / 24 Bottles

    Now, the sample can be filled into a separate bottle. 9. Discharge tube for not needed samples and rinsing water with external hose connection to return this water to the sewer. ba75817e01 05/2009 - 32 - PB 150/2 & PB 150-SE...
  • Page 33 The active program does not have to be interrupted and continues to run according to the programmed cycle, however, in the background. Detailed information can be found in the chapter „Programming“. Detailed view – distributor drive Detailed view – dosing/rinsing head ba75817e01 05/2009 - 33 - PB 150/2 & PB 150-SE...
  • Page 34: Sample Distribution At Pb 150-Se - Variants: 2 / 4 Bottles

    After expiration of the emptying time (adjustable), the respective rinsing valve is opened (rinsing time is adjustable, too) and the container is rinsed. After rinsing, the discharge valve closes and the device is ready for sampling again. ba75817e01 05/2009 - 34 - PB 150/2 & PB 150-SE...
  • Page 35: Troubleshooting - Distributor Pb 150-Se

    Check connections Distributor Program cannot be Wrong distributor has been Correct entry Programmin started entered Wrong positioning of Interfering pulses. Contact service Distributor distributor Pulse generator does not department function correctly ba75817e01 05/2009 - 35 - PB 150/2 & PB 150-SE...
  • Page 36: Programming

    According to the program section you are in, an action is triggered or, if a program is edited, you are automatically guided to the next menu. Remark: In general you can abort every action or step or return to a previous display by means of the back key. ba75817e01 05/2009 - 36 - PB 150/2 & PB 150-SE...
  • Page 37: Examples For Standard Programs

    A sample is e.g. extracted every 10 minutes. INTERVAL 00:10 ********************** Select whether the bottle filling shall be 5.2.2 ********************** BOTTLE FILLING time-dependent or depend on a pre-programmed number of samples *TIME-RELATED SAMPLE-RELATED ********************** ba75817e01 05/2009 - 37 - PB 150/2 & PB 150-SE...
  • Page 38 To start the program, press the „DOWN“ arrow key. ********************** 1.1.5 The just created program can either be • started immediately ********************** PROGRAM NO. 1 • started with a delay or CHANGE • changed DELAYED START *START ********************** ba75817e01 05/2009 - 38 - PB 150/2 & PB 150-SE...
  • Page 39 - At 20 mA 40 samples/hour shall be extracted. - If there are on an average 10 mA for a period of 1 hour, 20 samples will be extracted. ********************** ba75817e01 05/2009 - 39 - PB 150/2 & PB 150-SE...
  • Page 40 Remark: In that case, the program will run endlessly. So, ********************** please make sure that the bottles are emptied regularly according to the programming as otherwise they will be overfilled. ba75817e01 05/2009 - 40 - PB 150/2 & PB 150-SE...
  • Page 41 To start the program, press the „DOWN“ arrow key. 1.1.5 The just created program can either be • started immediately ********************** PROGRAM NO. 2 • started with a delay or • changed. CHANGE DELAYED START *START ********************** ba75817e01 05/2009 - 41 - PB 150/2 & PB 150-SE...
  • Page 42 A sample is e.g. extracted every 10 minutes. INTERVAL 00:10 ********************** Select whether the bottle filling shall be 5.2.2 ********************** BOTTLE FILLING 1. time-dependent or 2. depend on a preprogrammed number of samples. *TIME-RELATED SAMPLE-RELATED ********************** ba75817e01 05/2009 - 42 - PB 150/2 & PB 150-SE...
  • Page 43 To start the program, press the „DOWN“ arrow key. ********************** 1.1.5 The just created program can either be • started immediately ********************** PROGRAM NO. 3 • started with a delay or • changed. CHANGE DELAYED START *START ********************** ba75817e01 05/2009 - 43 - PB 150/2 & PB 150-SE...
  • Page 44: Main Menu (Root)

    • started with a delay or CHANGE DELAYED START • changed. *START ********************** MANUAL MODE 1.1.6 ********************** A single sample extraction into any bottle can be triggered MANUAL SAMPLE manually. EXTRACTION ********************** ba75817e01 05/2009 - 44 - PB 150/2 & PB 150-SE...
  • Page 45: Change Program

    3.4.1 TIME-related sampling ********************** By pressing the ENT key, the time-proportional sampling mode is MODE selected. EVENT *TIME FLOW ANAL. 0-20 mA ********************** ba75817e01 05/2009 - 45 - PB 150/2 & PB 150-SE...
  • Page 46 25. Select whether the bottle filling shall be time-dependent or whether 5.2.2 it shall depend on a pre-programmed number of samples. ********************** BOTTLE FILLING *TIME RELATED SAMPLE RELATED ********************** ba75817e01 05/2009 - 46 - PB 150/2 & PB 150-SE...
  • Page 47 Enter the max. filling time per bottle. PER BOTTLE 2:00H ********************** 5.2.5 ********************** Select your distributor/bottle variant. BOTTLE VERSION 12 x 1,0 L *12 x 2,9 L 24 x 0,4 L ********************** ba75817e01 05/2009 - 47 - PB 150/2 & PB 150-SE...
  • Page 48 Remark: In that case, the program will run endlessly. So, please make sure that the bottles are emptied regularly according to the programming as otherwise they will be ********************** overfilled. ba75817e01 05/2009 - 48 - PB 150/2 & PB 150-SE...
  • Page 49: Start Program

    ********************** 03.09.2003 17:28:17 SAMPLES TOTAL REQUESTED: TAKEN: ********************** Display of the taken and the requested samples of the current 3.3.4 bottle. ********************** 03.09.2003 17:28:18 SAMP. CURRENT BTL REQUESTED: TAKEN: ********************** ba75817e01 05/2009 - 49 - PB 150/2 & PB 150-SE...
  • Page 50 BOTTLE 1 This sample is stored in the memory as „manual sample“. ********************** ba75817e01 05/2009 - 50 - PB 150/2 & PB 150-SE...
  • Page 51 After expiration of this time, the system will PROGRAM NO. 1 return automatically to the running program. PAUSE (29:03) The pause mode can be left at any time by pressing the ENT key. *CONTINUE PROGRAM ********************** ba75817e01 05/2009 - 51 - PB 150/2 & PB 150-SE...
  • Page 52: Manual Mode

    ********************** ENT key again MANUAL SAMPLE enter a new bottle number for the extraction of the manual sample. BOTTLE: 1 ********************** ba75817e01 05/2009 - 52 - PB 150/2 & PB 150-SE...
  • Page 53: Setup

    2.8.4 ********************** SYSTEM SETTINGS Remark: Modifications should only be carried out by specially CHANGE trained service personnel. Wrong adjustments can lead to malfunctions or may even damage the apparatus! ********************** ba75817e01 05/2009 - 53 - PB 150/2 & PB 150-SE...
  • Page 54: Modem Settings

    4.15.1 ********************** If you entered a wrong number, press the left key to delete. NUMBER FOR SMS: Remark: Enter the country code without 0 or + ! 49174356984 > ********************** ba75817e01 05/2009 - 54 - PB 150/2 & PB 150-SE...
  • Page 55 ********************** program report (stored program data). SEND PROGRAM REPORT AUTOMAT.? ********************** 5.7.1 Possibility to send also the contents of the information memory. ********************** APPEND INFO MEMORY TO PROGRAM REPORT? ********************** ba75817e01 05/2009 - 55 - PB 150/2 & PB 150-SE...
  • Page 56 Operation 6.10.1 Select whether the info memory shall be deleted at the program ********************** start. DELETE INFO MEMORY AT PROGRAM START? ********************** ba75817e01 05/2009 - 56 - PB 150/2 & PB 150-SE...
  • Page 57: Change System Settings

    (utilizable signal if an optional relay is installed). However, the sampler continues to function even after a malfunction message. The program is not aborted for that reason! ba75817e01 05/2009 - 57 - PB 150/2 & PB 150-SE...
  • Page 58 Input the required pulse duration of the digital flow signal. According ********************** to our example, the signal would have to be present for 50 ms in PULSE DURATION order to be utilizable for the control. FLOW DIGITAL 50 [ms] ********************** ba75817e01 05/2009 - 58 - PB 150/2 & PB 150-SE...
  • Page 59: Extended Program Settings

    = fine adjustment (steps with 1). ENT=OK ESC=ABORT ********************** The „calibrated“ value is set immediately. Press 0 to reset to factor 1.0000. Confirm the values by pressing the ENT key! ba75817e01 05/2009 - 59 - PB 150/2 & PB 150-SE...
  • Page 60 7.5 °C section! ********************** 5.15.2 Set switch-off temperature of the cooling machine. ********************** COOLING OFF: Contact our service personnel before modifying the factory-set values in this section! 2.5 °C ********************** ba75817e01 05/2009 - 60 - PB 150/2 & PB 150-SE...
  • Page 61 (can be used as external signal). ********************** 4.17.27 Eight digital outputs are available. Each of them can be selected ********************** and configured. DIGITAL OUTPUTS *OUTPUT OUTPUT OUTPUT ********************** ba75817e01 05/2009 - 61 - PB 150/2 & PB 150-SE...
  • Page 62 0000 and confirm with ENT! 6299 Remark: In any case, note down the input code as otherwise ********************** you will have no more access to the protected program sections! ba75817e01 05/2009 - 62 - PB 150/2 & PB 150-SE...
  • Page 63: Memory

    -To see the oldest entry, press the left arrow key. 18.08.03 23-00 19.08.03 00-01 -To see the latest entry, press the right arrow key. 19.08.03 01-02 ---- 19.08.03 02-03 ---- You can scroll with the up/down keys. ********************** ba75817e01 05/2009 - 63 - PB 150/2 & PB 150-SE...
  • Page 64: Service Menu

    In the service menu the single components as well as the inputs ********************** and outputs can be tested. SERVICE MENU Remark: Modifications should only be carried out by specially trained service personnel. ********************** ba75817e01 05/2009 - 64 - PB 150/2 & PB 150-SE...
  • Page 65 3.18.15 With this function a system check is possible. ********************** SELF-TEST Remark: A system check must only be carried out by specially trained service personnel! DISPLAY PROTOCOL *START TEST ********************** ba75817e01 05/2009 - 65 - PB 150/2 & PB 150-SE...
  • Page 66 • EVENT 0 (1) 0 = There is no event signal 1 = There is an event signal • FLOW (digital) 0 (1) 0 = No pulse 1 = Pulse (digital) ba75817e01 05/2009 - 66 - PB 150/2 & PB 150-SE...
  • Page 67: Appendix A - Special Programming

    A detailed description (example) can be found on the next page. 7.1.1 Enter the time at which the sampler shall change to bottle No. 1. ********************** START TIME FIRST BOTTLE 00:00 ********************** ba75817e01 05/2009 - 67 - PB 150/2 & PB 150-SE...
  • Page 68 This entry enables the SAMPLES extraction of a big sample volume within a short time. ********************** 5.2.7 Select on which bottle the program shall be started. ********************** FIRST BOTTLE ********************** ba75817e01 05/2009 - 68 - PB 150/2 & PB 150-SE...
  • Page 69 The sampler extracts samples according to these parameters until the event signal is released. Then the sampler continues to carry 01:00H out sample extractions according to the previously selected ********************** sampling mode! ba75817e01 05/2009 - 69 - PB 150/2 & PB 150-SE...
  • Page 70 Possibility to have a further special days´ setting for example for ********************** holidays etc. The programming procedure is the same as at SPECIAL DAYS 2 „Special days 1“. *SUNDAY MONDAY TUESDAY ********************** ba75817e01 05/2009 - 70 - PB 150/2 & PB 150-SE...
  • Page 71: Special Display/Malfunction Messages

    „right“ bottle. 4. Error DISTRIBUTOR This message is displayed if there is a malfunction at the ******************** distributor! PROGRAM NO. 02 START 12.03.02/08:00 ERROR DISTRIBUTOR PAUSE WITH ENT ******************** ba75817e01 05/2009 - 71 - PB 150/2 & PB 150-SE...
  • Page 72: Extended Malfunction Messages

    VOLTAGE LOSS UNTIL 4.23 SAMP. TAKEN (BTL) 4.24 SAMP. REQ. (BTL) 4.25 ERROR ANAL. SIG FROM 4.26 ERROR ANAL. SIG ´TIL 4.28 SYSTEM START 4.29 MAX. FILLING TIME 4.31 INTERNAL TEMPERATURE ba75817e01 05/2009 - 72 - PB 150/2 & PB 150-SE...
  • Page 73: Digital Inputs / Outputs

    Error: ANALOGUE SIGNAL 14.13 Error: VOLTAGE LOSS 14.16 Counter: SAMPLING OK 14.17 Counter: SAMPLING ERROR 14.18 Counter: DIGITAL PULSE DIGITAL INPUTS: 15.2 START PROGRAM 15.3 STOP PROGRAM 15.4 SAMPLING 15.5 BOTTLE CHANGE ba75817e01 05/2009 - 73 - PB 150/2 & PB 150-SE...
  • Page 74 Possibility to define a minimum flow rate for the sample extraction. ********************** If the flow rate is lower than the set value, there is no sample MIN. FLOW FOR extraction! SAMPLE EXTRACTION 000.0 mA ********************** ba75817e01 05/2009 - 74 - PB 150/2 & PB 150-SE...
  • Page 75 *START Press the ENT key to start the calibration. ********************** ba75817e01 05/2009 - 75 - PB 150/2 & PB 150-SE...
  • Page 76 ********************** here and confirmed by pressing the ENT key. SAMPLE VOLUME Press the ESC key to leave the calibration mode! 150 ml Now the system is ready for operation! ********************** ba75817e01 05/2009 - 76 - PB 150/2 & PB 150-SE...
  • Page 77 2.0 s ********************** 4.17.13 Adjust the dosing pinch valve power in %. ********************** PINCH-VALVE Modifications should only be carried out by (DOSING) POWER specially trained service personnel! 40 % ********************** ba75817e01 05/2009 - 77 - PB 150/2 & PB 150-SE...
  • Page 78 *OFF Remark: According to the selected direction, the pinch valve ********************** always moves up to the limit stop, that means it has to be stopped manually in the desired position!!! ba75817e01 05/2009 - 78 - PB 150/2 & PB 150-SE...
  • Page 79 TAKE OUT SAMPLE BOTTLE 1 ********************** 6.1.1 This menu is shown while the bottle is moved on the sample taking- ********************** out position. BOTTLE IS MOVED ON SAMPLE TAKING –OUT POS. ********************** ba75817e01 05/2009 - 79 - PB 150/2 & PB 150-SE...
  • Page 80 1. Select the bottle number already displayed by pressing TAKE OUT SAMPLE the ENT key again 2. Enter the bottle from which the sample shall be collected. BOTTLE: 1 ********************** ba75817e01 05/2009 - 80 - PB 150/2 & PB 150-SE...
  • Page 81 The number of the bottle emptied/rinsed at the moment is shown on EMPTYING BOTTLE: 1 the display. You can abort this action at any time by pressing the ESC key. ESC TO CANCEL ********************** ba75817e01 05/2009 - 81 - PB 150/2 & PB 150-SE...
  • Page 82 1= Sensor inactive P= Bottle position *OFF P1=Bottle 1 ********************** V=Bottle valve Same function as described above, however, for the bottle valve 3.18.9 motor. ********************** BOTTLE VALVE MOTOR [P=0 P1=0 V=0] ********************** ba75817e01 05/2009 - 82 - PB 150/2 & PB 150-SE...
  • Page 83 [P=0 P1=0 V=0] OPEN *CLOSED ********************** 3.18.11 Possibility to move the distributor on any bottle position in order to ********************** effect a position control. DISTRIBUTOR 12 BOTTLES BOTTLE: 1 ********************** ba75817e01 05/2009 - 83 - PB 150/2 & PB 150-SE...
  • Page 84 5. Connect the ends of the Y-connectors with the pump according to the diagram: 2=pressure, 4=vacuum, 5=metering vessel and 3=air filter. 6. Check the correct functioning of the valve system by triggering a manual sample. ba75817e01 05/2009 - 84 - PB 150/2 & PB 150-SE...
  • Page 85 Unscrew the 4 screws, remove the pump head (1) and replace it. Proceed in the same way to replace the diaphragm (2). No. 1 Pump head Art. No. BM60043 No. 2 Diaphragm Art. No. BM60042 ba75817e01 05/2009 - 85 - PB 150/2 & PB 150-SE...
  • Page 86 The adjusted pumping capacity has no effect on the suction height but only on the average suction speed when extracting the sample. The lower the pumping capacity, the longer the life of the pump! ba75817e01 05/2009 - 86 - PB 150/2 & PB 150-SE...
  • Page 87: Spare Parts / Accessories Pb 150/2

    Spare parts, accessories, circuit diagrams Spare parts / accessories PB 150/2 Spare parts and accessories for your sampler, see the current price list for the catalog „On-line instrumentation“. ba75817e01 05/2009 - 87 - PB 150/2 & PB 150-SE...
  • Page 88: Circuit Diagram Mains Connection Pb 150/2 & Pb150-Se

    Spare parts, accessories, circuit diagrams Circuit diagram mains connection PB 150/2 & PB150-SE Motor 4: Cooling machine Motor 5: Circulation fan Motor 6: Exhaust fan ba75817e01 05/2009 - 88 - PB 150/2 & PB 150-SE...
  • Page 89: Circuit Diagram Vacuum System

    Spare parts, accessories, circuit diagrams Circuit diagram vacuum system Motor 1: Diaphragm pump Motor 2: Valve system Motor 3: Pinch valve Motor 7: Distributor drive ba75817e01 05/2009 - 89 - PB 150/2 & PB 150-SE...
  • Page 90: Circuit Diagram Bypass System

    Spare parts, accessories, circuit diagrams Circuit diagram bypass system Motor 2: Dosing pinch valve Motor 3: Pinch valve Motor 7: Distributor drive ba75817e01 05/2009 - 90 - PB 150/2 & PB 150-SE...
  • Page 91: Circuit Diagram Flow-Proportional System

    Motor 1: Big diaphragm pump Motor 8: Small diaphragm pump Motor 2: Valve system Sensor 5: Pulse transmitter Motor 3: Pinch valve MV 1: Aeration valve Motor 7: Rotating distributor ba75817e01 05/2009 - 91 - PB 150/2 & PB 150-SE...
  • Page 92: Circuit Diagram Terminal Connections

    Spare parts, accessories, circuit diagrams Circuit diagram terminal connections Eingangssignale / Input signals / Signaux d´entrée analog /analogue /analogique + analog /analogue /analogique - digital/digital/numérique Ereignis/event/événement ba75817e01 05/2009 - 92 - PB 150/2 & PB 150-SE...
  • Page 93 Spare parts, accessories, circuit diagrams Circuit diagram self-emptying PB 150/2 & PB 150-SE 12 x 1,6 Motor 1: Diaphragm pump Sensor 5: Position Output terminal 6 Motor 2: Valve system Sensor 6: Position1 Output terminal 5 Motor 3: Pinch valve...
  • Page 94: Circuit Diagram Self-Emptying Pb 150/2-A 12 X 1,6

    Spare parts, accessories, circuit diagrams Circuit diagram self-emptying PB 150/2 & PB 150-SE 24 x 2 L Motor 1: Diaphragm pump Sensor 5: Position Output terminal 6 Motor 2: Valve system Sensor 6: Position1 Output terminal 5 Motor 3: Pinch valve...
  • Page 95: Circuit Diagram Pb 150/2-A 12 X 1,6 Bypass

    Spare parts, accessories, circuit diagrams Circuit diagram PB 150/2 & PB 150-SE 12 x 1,6 Bypass Motor 2: Dosing pinch vale Sensor 5: Position Output terminal 6 Motor 3: Pinch valve Sensor 6: Position1 Output terminal 5 Motor 7: Distributor drive...
  • Page 96: Circuit Diagram Pb 150/2-A 24 X 2,0 Bypass

    Spare parts, accessories, circuit diagrams Circuit diagram PB 150/2 & PB 150-SE 24 x 2,0 Bypass Motor 2: Dosing pinch vale Sensor 5: Position Output terminal 6 Motor 3: Pinch valve Sensor 6: Position1 Output terminal 5 Motor 7: Distributor drive...
  • Page 97 Spare parts, accessories, circuit diagrams Circuit diagram PB 150/2 & PB 150-SE 12 x 1,6 L flow-proportional Motor 1: Diaphragm pump Sensor 5: Position Output terminal 6 Motor 2: Valve system Sensor 6: Position1 Output terminal 5 Motor 3: Pinch valve...
  • Page 98: Circuit Diagram Pb 150/2-A 12 X 1,6 L Flow-Proportional

    Spare parts, accessories, circuit diagrams Circuit diagram PB 150/2 & PB 150-SE 12 x 1,6 L flow-proportional Motor 1: Diaphragm pump Sensor 5: Position Output terminal 6 Motor 2: Valve system Sensor 6: Position1 Output terminal 5 Motor 3: Pinch valve...
  • Page 99: Circuit Diagram Pb 150-Se 2 X 10 L Vacuum

    Circuit diagram PB 150-SE 2 x 10 L vacuum Motor 1: Diaphragm pump Motor 2: Valve system Motor 3: Pinch valve Motor 6: Distributor drive Motor 7: Bottle valve drive 1 Motor 8: Bottle valve drive 2 ba75817e01 05/2009 - 99 - PB 150/2 & PB 150-SE...
  • Page 100: Circuit Diagram Optional Inlet / Additional Valve 900276

    Spare parts, accessories, circuit diagrams Circuit diagram optional inlet / additional valve 900276 ba75817e01 05/2009 - 100 - PB 150/2 & PB 150-SE...
  • Page 101: Circuit Diagram Optional Inlet / Additional Valve 900276

    Spare parts, accessories, circuit diagrams Circuit diagram optional inlet / additional valve 900276 ba75817e01 05/2009 - 101 - PB 150/2 & PB 150-SE...
  • Page 102: Warranty And Liability

    Any further claims, in particular claims for consequential damages cannot be made. Consumables and damage caused by improper handling, poor installation or incorrect use are excluded from this clause. ba75817e01 05/2009 - 102 - PB 150/2 & PB 150-SE...
  • Page 103 (Electrical Equipment for measurement, control and laboratory use—EMC requirements), class “A” emission limits. Supporting test records by MAXX GmbH, certified compliance by MAXX GmbH. Additional Emission Standards include: EN 55011:08.03, Class “A” emission limits. ba75817e01 05/2009 - 103 - PB 150/2 & PB 150-SE...
  • Page 106 Wissenschaftlich-Technische Werkstätten GmbH Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim Germany Tel: +49 (0) 881 183-0 +49 (0) 881 183-100 Fax: +49 (0) 881 183-420 E-Mail: Info@WTW.com Internet: http://www.WTW.com...