wtw inoLab Level 3 Operating Manual page 187

Measuring system with active multifunction boxes
Table of Contents

Advertisement

Accepting the cell
constant of the
calibration record
Selecting any cell
constant
Accept the cell constant of the last calibration record as
follows:
1
Switch to the Menu mode.
46
2
Use
3
Select the Cell const. menu item with
4
Confirm the Cell const. menu item.
5
Select the Cal cell menu item with
6
Confirm the Cal cell menu item with
The cell constant of the last calibration is accepted.
Cal cell is active (✔).
7
Terminate the Menu mode with
You can set the cell constant manually in the range 0.250 ...
-1
2.500 cm
as follows:
1
Switch to the Menu mode.
46
2
Use
3
Select the Cell const. menu item with
4
Confirm the Cell const. menu item.
5
Select the Man cell menu item with
to select the main Cal menu.
to select the main Cal menu.
Conductivity
82
82
.
g
.
m
.
82
82
.
187