wtw inoLab Level 3 Operating Manual page 86

Measuring system with active multifunction boxes
Table of Contents

Advertisement

pH
ConCal
single-point
calibration
ConCal
two-point calibration
86
Only one buffer solution is required for single-point
calibration. The pH value should lie as close as possible to
the expected measured value. Only the electrode
asymmetry is then newly determined. The slope of the last
two-point calibration is retained.
Use the following buffer solutions:
l
pH 7.0 ± 0.5
l
Any other buffer solution so that the measuring range is
covered as well as possible
Note
The ConCal calibration procedure has no temperature
compensation. The buffer solutions must have the correct
temperature.
If you use a temperature probe, it is not necessary to enter
the temperature manually.
1
Select the ConCal calibration procedure
(section 6.2.1).
2
Perform the preparatory activities (see section 6.2.2).
3
Start the calibration procedure with
4
Immerse the electrode and the temperature probe in
the pH 7.0 ± 0.5 buffer solution.
5
If necessary, set the temperature of the first buffer
solution with
6
Use [Adj. Asy.] to call up the input field for the nominal
pH value of the buffer solution.
82
7
Use
(and
of the buffer solution (at the current temperature).
8
Confirm the nominal pH value with
9
Use [ESC] to leave the input field for the nominal pH
value.
82
.
46
) to select the nominal pH value
c
.
g
.