Routines; Setting Up Routines - wtw inoLab Level 3 Operating Manual

Measuring system with active multifunction boxes
Table of Contents

Advertisement

BOD
170
9.4

Routines

9.4.1

Setting up routines

Routines define constantly recurring measurement
procedures. This includes the number and name of blank
value solutions and sample dilutions as well as the dilution
ratios.
Routines are set up
l
Before the first measurement:
Complete routines can be set up in the BOD menu.
l
At the end of a start measurement:
You can save the current settings as a routine in the Save
routine window.
Have the following data ready before setting up a routine:
l
Number of blank value solutions (max. 5)
l
Number of sample dilutions per sample (max. 18)
l
Amount of sample (in ml) and total volume (in ml) of each
sample dilution
l
Incubation period (5 to 30 days)
l
Names of the blank value solutions, samples, sample
dilutions and routines (max. of 12 characters each)