Ion concentration
106
82
8
Use
to enter the concentration of standard
solution 2.
9
Thoroughly rinse the electrode with deionized water.
10
Immerse the electrode in standard solution 2.
g
11
Press the
The electrode voltage is displayed and the
display indicator flashes.
12
As soon as a stable value is reached,
h
stops flashing. The slope (mV) of the electrode is
displayed after the two-point calibration. The probe
symbol shows the electrode evaluation after the two-
point calibration.
key.
h