wtw inoLab Level 3 Operating Manual page 85

Measuring system with active multifunction boxes
Table of Contents

Advertisement

g
21
Press the
h
The
display indicator flashes.
The electrode voltage (mV) appears on the display.
As soon as a stable value is reached,
The probe symbol shows the calibration evaluation
after the four-point calibration.
The value of the slope S (in mV/pH) and the value of
the asymmetry U
g
22
Press the
The pH value and the probe symbol appear on the
display.
The system is calibrated for pH
measurements.
Note
The calibration line is determined by linear regression.
After the calibration is finished, the calibration record is
automatically printed if the printer is ready for operation.
Information on setting up a printer and on the printout of
further calibration protocols is given in chapter 12 D
and section 12.2.4 P
OUTPUT
key.
(in mV) appear on the display.
Asy
key.
.
RINTING
pH
h
disappears.
ATA
85