wtw inoLab Level 3 Operating Manual page 79

Measuring system with active multifunction boxes
Table of Contents

Advertisement

Selecting the
calibration
procedure
Selecting the buffer
data record
1
Switch to the Menu mode.
46
2
Use
3
Select the pH caltype menu item with
4
Confirm the pH caltype menu item with
82
5
Use
to select the AutoCal or ConCal menu item.
6
Confirm the selected menu item with
The selected calibration procedure is active (✔).
For each calibration procedure, you can select between the
buffer datasets TEC, DIN and Mrk:
7
Select the pH Buffer menu item with
8
Confirm the pH Buffer menu item with
9
Select the Buffer TEC, Buffer DIN or Buffer Mrk 1...3
menu item with
10
Confirm the selected buffer data record with
selected setting is active (✔).
11
Terminate the Menu mode with
to select the main Cal menu.